Control Computer/Software
A state of the art PC, optimized for AFM operation, is included with the Nano-DST™ system. Reliability is assured with computers configured by Dell™. The computer system includes ample disk drive capacity to store AFM images and to transfer them to a CD when desired. A 19” LCD monitor is also included.
The Nano-DST™ command and control software, derived from SPM Cockpit™, is in use in hundreds of laboratories throughout the world. The software operates in two modes:
• EZ Mode™ for beginners and casual users
• X’pert™ Mode for advanced AFM operators
Signals may be monitored with the Signal Access Console that connects to the rear of the electronic
controller. All signals associated with the performance of the Nano-DST™ are available.
Environmental
Vibration Solution: The highest performance is assured when the optional acoustic enclosure and vibration table are used with the Nano-DST™.
System Options:
• STM
• Environmental Cell
• SHARK
• MFM
• EFM
• Kelvin Probe
System Options and Specifications
Light Lever AFM / Scanner
| Force Sensor |
Light Lever |
| Detector |
4 Quadrant Photo-detector |
| Laser |
650 nm diode – 25 micron spot size |
| Laser Adjustment |
80 TPI |
| Probe Holder |
Universal Adapter |
| Probe Types |
Pre-mounted - Magnetic |
| XYZ Calibration Sensors |
Inductive – 1 nm resolution |
| Z Scanner Type |
Piezoelectric Ceramic – Direct |
| Z Range |
8 Microns |
| Z Noise |
<0.06 nm* |
| Z Linearity |
< 1 % |
| Z out of plane motion |
< +/- 30 nm |
| XY Scanner Type |
Flexure |
| XY Range |
90 Microns(180 or 360 microns optional) |
| XY Linearity |
< 1 % |
| XY Noise |
< 0.01 nm (open loop) |
| XY, XZ, YZ Crosstalk |
< 1 % |
| Probe Electrical Connector |
Isolated 5 Connector |
1624 Electronic Control
| Primary Scan DAC |
24 Bits |
| Secondary Scan DAC |
16 Bits |
XYZ board – Primary and Secondary
• Digital Zoom/Offset
• Analog Zoom
• Z High Voltage
• Z Sample/Hold
• PID Control
• Phase/Amplitude
• XY High Voltage
• XY Sensor Feedback |
16 Bits
4 bits on High Voltage
-20 to 140 Volts
20 Bits
Analog
400 Mhz Clock/32 Bit Frequency Control
+X,-X +Y,-Y (-20 to 140 Volts)
Analog PID Control |
| User Access |
Connector for Signal Access Console |
Internal Digital Control
• CPU
• Operating System
• Communications Protocol
• Memory |
Pentium
Linux
TCP/IP
RAM/Disk Drive |
Motor Driver Board
• Number of Motor Drivers
• Motor Type
• Control
• Additional Function |
7
Stepper
Microprocessor
Light Intensity Control |
Stage
XY Stage Range
• X
• Y |
50 mm
75 mm |
| XY Stage Step Size |
2 um |
| XY Stage Accuracy |
< 10 um |
| Z Motor Range |
8 mm |
| Z Motor Step Size |
162 nm |
Sample Puck
• Hold Down
• Height
• Maximum Sample Diameter |
Magnetic
30 mm
100 mm |
Optical Microscope
• Resolution
• Zoom Ratio
• Field of View |
1.5 u
4:1
140 ? 190 u at Maximum Zoom |
| Probe Exchange |
Flip Up |
| Physical Specifications |
| Width |
Depth |
Height |
Weight (kg) |
|
• Stage
• Vibration Enclosure
• Electronic Controller
• Work Station |
| 14 |
19 |
16 |
50 |
| 31 |
33 |
56 |
275 |
| 12 |
18 |
22 |
22 |
| 7 |
16 |
16 |
14 |
|
Rapid Scanner
| Type |
Four Quadrant Tube |
| XY Range |
2 Microns |
| XY Noise |
0.1 nm |
| Z Range |
500 nm |
| Z Noise |
0.001nm |
| Scan Rate |
250 Hz |
Control Station
| Computer Monitor |
19” LCD |
| Video Monitor |
19” LCD |
| Accessories |
Mouse, Keyboard, Trackball |
Computer:
• Type
• Operating System
• Memory
• Disk Drive
AFM Software:
• Acquisition
• Display
Video Converter |
IBM PC
Microsoft Vista
> 1 Gigabyte
> 200 G Bytes
EZ Mode™, X’ert™ Mode
AFM Analysis, and NanoRule+™
NTSC – SVGA |
Vibration Control (Recommended)
| Acoustic Shield |
Herzan AEK-2002 Enclosure |
| Vibration Table |
Herzan TS-150 Active Isolation System |
* Vibration environment can lower noise preformance.